Fluke
Mostrando 73–81 de 696 resultadosOrdenado por los últimos
-
Medición de Temperatura
Calibrador de temperatura Fluke 9190A-B-P-156
DRY-WELL, ULTRA-COLD, INSRT-B, BLT-IN ELCT, 115V
- Wide temperature range
- –95 °C to 140 °C
- Excellent accuracy
- Accuracy using built-in reference thermometer readout: ± 0.05 °C full range
- Display accuracy: ± 0.2 °C full range
- Best-in-class stability
- ± 0.015 °C full range
- Fast cooling time
- 23 °C to –90 °C: 80 minutes
- 23 °C to –95 °C: 90 minutes
- 140 °C to 23 °C: 60 minutes
- Portability
- Weighs only 16 kg (35 lbs)
- Built-in front and back handles for easy two-handed carry
- Best measurement practices
- Conforms with EURAMET cg-13 guidance on measurement practices for temperature calibrators
- Ergonomics
- Two-handed carry; built-in handles on front and back
- All controls and inputs on the front panel
- Stability indicator
- Shows whether block temperature is stable and within limits defined by the user
- PC interface
- USB and RS-232 support remote operation from a computer
- Dual heater zone
- Active heater zone control compensates for differential temperatures between zones
- Minimizes axial temperature gradients.
(0 reviews) - Wide temperature range
-
Medición de Temperatura
Pozo de metrología Fluke 9190A-B-156
DRY-WELL, ULTRA-COLD, INSRT-B, 115V
- Wide temperature range
- –95 °C to 140 °C
- Excellent accuracy
- Accuracy using built-in reference thermometer readout: ± 0.05 °C full range
- Display accuracy: ± 0.2 °C full range
- Best-in-class stability
- ± 0.015 °C full range
- Fast cooling time
- 23 °C to –90 °C: 80 minutes
- 23 °C to –95 °C: 90 minutes
- 140 °C to 23 °C: 60 minutes
- Portability
- Weighs only 16 kg (35 lbs)
- Built-in front and back handles for easy two-handed carry
- Best measurement practices
- Conforms with EURAMET cg-13 guidance on measurement practices for temperature calibrators
- Ergonomics
- Two-handed carry; built-in handles on front and back
- All controls and inputs on the front panel
- Stability indicator
- Shows whether block temperature is stable and within limits defined by the user
- PC interface
- USB and RS-232 support remote operation from a computer
- Dual heater zone
- Active heater zone control compensates for differential temperatures between zones
(0 reviews) - Wide temperature range
-
Medición de Temperatura
Calibrador de temperatura Fluke 9190A-A-P-156
DRY-WELL, ULTRA-COLD, INSRT-A, BLT-IN ELCT, 115V
- Wide temperature range
- –95 °C to 140 °C
- Excellent accuracy
- Accuracy using built-in reference thermometer readout: ± 0.05 °C full range
- Display accuracy: ± 0.2 °C full range
- Best-in-class stability
- ± 0.015 °C full range
- Fast cooling time
- 23 °C to –90 °C: 80 minutes
- 23 °C to –95 °C: 90 minutes
- 140 °C to 23 °C: 60 minutes
- Portability
- Weighs only 16 kg (35 lbs)
- Built-in front and back handles for easy two-handed carry
- Best measurement practices
- Conforms with EURAMET cg-13 guidance on measurement practices for temperature calibrators
- Ergonomics
- Two-handed carry; built-in handles on front and back
- All controls and inputs on the front panel
- Stability indicator
- Shows whether block temperature is stable and within limits defined by the user
- PC interface
- USB and RS-232 support remote operation from a computer.
(0 reviews) - Wide temperature range
-
Medición de Temperatura
Pozo de metrología Fluke 9190A-A-156
DRY-WELL, ULTRA-COLD, INSRT-A, 115V
- Wide temperature range
- –95 °C to 140 °C
- Excellent accuracy
- Accuracy using built-in reference thermometer readout: ± 0.05 °C full range
- Display accuracy: ± 0.2 °C full range
- Best-in-class stability
- ± 0.015 °C full range
- Fast cooling time
- 23 °C to –90 °C: 80 minutes
- 23 °C to –95 °C: 90 minutes
- 140 °C to 23 °C: 60 minutes
- Portability
-
- Weighs only 16 kg (35 lbs)
- Built-in front and back handles for easy two-handed carry
- Best measurement practices
- Conforms with EURAMET cg-13 guidance on measurement practices for temperature calibrators
- Ergonomics
- Two-handed carry; built-in handles on front and back
- All controls and inputs on the front panel
- Stability indicator
- Shows whether block temperature is stable and within limits defined by the user
- PC interface
- USB and RS-232 support remote operation from a computer
- Dual heater zone
- Active heater zone control compensates for differential temperatures between zones
- Minimizes axial temperature gradients
- Ramp-and-soak set-points
- Define up to eight set-point temperatures; automatical.
(0 reviews) - Wide temperature range
-
Medición de Temperatura
Pozo de metrología Fluke 9173-F-R-156
METROLOGY WELL, 700C, W/9173-INSF, W/REF, 115V
- Best-performing industrial heat sources (accuracy, stability, uniformity) in the world
- Immersion depth to 203 mm (8 in)
- Optional ITS-90 reference input reads PRTs to ±0.006 °C
- Temperature range from –45 °C to 700 °C.
(0 reviews) -
Medición de Temperatura
Pozo de metrología Fluke 9173-F-156
METROLOGY WELL, 700C, W/9173-INSF, 115V
- Immersion depth to 203 mm (8 in)
- Best-performing industrial heat sources (accuracy, stability, uniformity) in the world
- Optional ITS-90 reference input reads PRTs to ±0.006 °C
- Temperature range from –45 °C to 700 °C.
(0 reviews) -
Medición de Temperatura
Pozo de metrología Fluke 9173-E-R-156
METROLOGY WELL, 700C, W/9173-INSE, W/REF, 115V
- Best-performing industrial heat sources (accuracy, stability, uniformity) in the world
- Immersion depth to 203 mm (8 in)
- Optional ITS-90 reference input reads PRTs to ±0.006 °C
- Temperature range from –45 °C to 700 °C.
(0 reviews) -
Medición de Temperatura
Pozo de metrología Fluke 9173-E-156
METROLOGY WELL, 700C, W/9173-INSE, 115V
- Immersion depth to 203 mm (8 in)
- Best-performing industrial heat sources (accuracy, stability, uniformity) in the world
- Optional ITS-90 reference input reads PRTs to ±0.006 °C
- Temperature range from –45 °C to 700 °C.
(0 reviews) -
Medición de Temperatura
Pozo de metrología Fluke 9173-DW-R-156
METROLOGY WELL,700C,NO INSRT,W/BLTIN REF,115V
- Best-performing industrial heat sources (accuracy, stability, uniformity) in the world
- Immersion depth to 203 mm (8 in)
- Optional ITS-90 reference input reads PRTs to ±0.006 °C
- Temperature range from –45 °C to 700 °C.
(0 reviews)