Descripción
High Accuracy Dual Channel Thermistor Probe Wifi Temperature Data Logger
Now Cloud enabled!
The EL-WiFi-DTP+ data logger measures the temperature of the environment in which the two probes are situated. This high accuracy sensor is typically accurate to ±0.2°C (-15°C to +80°C). The unit is freestanding, but can be attached to a wall or surface using the bracket provided. The unit can be clipped in and out of the bracket as required. The probes are connected to the sensor via a 160mm Y splitter cable which is provided.
Data is streamed wirelessly to a local PC or to our Cloud service. During configuration, the data logger will search for an existing wireless network whilst physically connected to the PC. It can then be placed anywhere within range of the network. If the data logger temporarily loses connectivity with the network, it will log readings until the connection is re-established (max 30 days at 10 second sample interval).
Our sensors require your wireless network to be IEEE 802.11b compliant. If you have an enterprise network or multiple access points with the same SSID, please contact us to discuss your application before placing your order.
Calibration Certificates
Lascar’s calibration certificates are traceable to UKAS or NIST standards depending on your location. Customers in the Americas, or customers ordering through our US sales office will receive NIST traceable certificates generated in our ISO 17025 facility, whilst customers in the UK, EU, Rest-of-world or customers ordering through our UK headquarters will receive UKAS traceable certificates.
A calibration certificate is not supplied as standard with a data logger.
Please note that a calibration certificate can only be provided with a data logger at the time of original purchase.
Manufacturer Specs
| Specifications | Minimum | Typical | Maximum | Unit |
| Battery life | >6* | Months | ||
| USB supply voltage | 4.5 | 5.5 | Vd.c. | |
| Operating temperature range (sensor only) | -20 (-4) | +60 (+140) | °C (°F) | |
| Logging period (user configurable) | 10 sec | 10 min | 12 hrs | |
| Transmission period (user configurable) | 1 min | 1 hr | 24 hrs | |
| Temperature measurement range | -40 (-40) | +125 (+257) | °C (°F) | |
| Temperature measurement resolution | 0.01 | °C | ||
| Temperature display resolution | 0.01 | °C | ||
| Temperature accuracy (probe dependent) | ±0.2°C (-15 to +80°C) | ±0.6°C (-40 to +125°C) | °C |
Warning – do not exceed operating temperatures




Valoraciones
No hay valoraciones aún.