Descripción

Comet U0111
- Temperature range: (-90 to +260) °C, Pt1000/3850 ppm sensor
- Measuring current: ca 0.5 mA with pulses of ca 60 ms length
- High quality, accurate and stable sensors
- Built-in battery operated
Temperature Data Logger for One External Pt1000 Probe
Datalogger is designed to record temperature from one external Pt1000 probe. The recording is performed in non-volatile electronic memory. The data can be transferred to a PC via USB-C.
In case of exceeded set limits alarms are indicated by LED, LCD, and acoustically by built-in beeper. The device includes a Traceable calibration certificate with declared metrological traceability of etalons is based on requirements of EN ISO/IEC 17025 standard.
This model can measure one, two, or up to four temperature values transmitted by Pt1000/E series external probes. The response to a temperature jump depends on the connected probe design. As a rule, this model is a few times faster than that using the inner sensor. It is often used to monitor locations where the device itself is not directly in the measuring space and only probes are in. The maximum length of each probe lead should not exceed 15 m. It is recommended to use shielded cables. To maintain the IP rating, the unused probe connectors should be provided with a supplied closing cap. The battery in the device cannot be charged.

Manufacturer Specs
| Technical Data | |
| Measuring range | 90 to +260 °C |
| Accuracy | ±0.2°C; ±0,2 % of the measurement value in a range of +100 to +260 °C |
| Resolution | 0.1 °C |
| Response time t90 of temperature measurement (temperature step 20°C, air flow approximately 1m/s) | according to the connected probe |
| General Technical Data | |
| Operating temperature | -30 to +70 °C |
| Channels | 1x external temperature probe |
| Memory | 500,000 values in noncyclic logging mode; 350,000 values in cyclic record mode |
| Recording interval | adjustable from 1 s to 24 h |
| Display and alarm refresh | adjustable 1 s, 10 s, 1 min |
| Recording mode | noncyclic – data logging stops after filling the memory cyclic – after filling memory oldest data is overwritten by new |
| Real time clock | year, leap year, month, day, hour, minute, second |
| Power | Lithium battery 3.6 V, size AA |
| Protection class | IP67 |
| Dimensions | 61 x 93 x 32 mm |
| Weight (including batteries) | approx. 130 g |
| Warranty | 3 years |
Tech Specs
Dataloggers / Data Acquisition/Dataloggers for Environmental Variables/Temperature Dataloggers Templ |
|
|---|---|
| Max Temperature | 260 C |
| Min Temperature | -90 C (-130 F) |
| Power Source | Battery |
| # Inputs/Channels | 1 |
| SAMPLE RATE/Intervals | Adjustable from 1s to 24h |
| Accuracy | ±0.2°C; ±0,2 % of the measurement value in a range of +100 to +260 °C |
| LCD Screen | Yes |
| Software Included | Yes |
Test Equipment General Attributes |
|
| Warranty | 3 YEARS |
| IP Rating | IP67 |
| Product Weight | 0.29 LBS |
| Shipping Weight | 0.33 LBS |
| Battery Type | Li-ion |
| Power Supply Voltage | Battery Powered |
| Country of Origin | Czech Republic |




Valoraciones
No hay valoraciones aún.