Descripción

Comet U3631
- A solution for every need, every budget – economy and premium dataloggers
- High quality, accurate and stable sensors
- Built-in battery operated
- Built-in printer
Temperature and Humidity Data Logger with Connector for Another Temperature Probe
Datalogger is designed to record temperature and humidity from built-in sensors and from one external temperature probe Pt1000. The recording is performed in non-volatile electronic memory. The data can be transferred to a PC via USB-C.
In case of exceeded set limits alarms are indicated by LED, LCD, and acoustically by built-in beeper. Recorder includes a Traceable calibration certificate with declared metrological traceability of etalons is based on requirements of EN ISO/IEC 17025 standard.

Compact thermometer – hygrometer with optional external temperature probe
This model is designed to measure temperature, relative humidity, and dew-point temperature by internal sensor, moreover, one additional external temperature probe Pt1000/E can be connected. Internal sensors are located under the grid on the front of the unit and are protected with Teflon foil. Foil permits water vapour but prevents water from permeating the liquid state. The device also measures the temperature from an external probe and the difference between this temperature and the dew point temperature.
If the surface temperature of the material is measured by an external probe, the risk of condensation of water vapor on the surface to be measured can be determined immediately. The device alone is suitable for measuring in places where there is no rapid change in temperature or relative humidity, and there is no condensation of water vapours. If water vapour condensation occurs inside the datalogger, then the resulting water will remain there and may damage its electronics. The device is placed directly in the measured space. The battery in the device cannot be charged. It is recommended to calibrate this device in the laboratory with the unscrewed bottom lid of the box to allow the air to flow from the back of the electronics. When airflow in the range of 0.5 – 1 m/s perform reading after about 4 hours.
The closed device reaches a difference of approximately 2 %RH when inserted in the humidity chamber, then dramatically slows down. Precise reading takes many times longer time than with the open rear lid. The maximum length of each probe lead should not exceed 15 m. It is recommended to use shielded cables. To maintain the IP rating, the unused probe connector should be provided with a supplied closing cap.
Manufacturer Specs
| TEMPERATURE SENSOR | |
| Measuring range | -30 to +70 °C |
| Accuracy | ±0.4 °C |
| Resolution | 0.1 °C |
| TEMPERATURE SENSOR – external probe Pt1000 | |
| Measuring range | -90 to +260 °C |
| Accuracy | ±0.2 °C (-50 to +100 °C); ±0.2 % from reading from +100 to +260 °C; ±0.4 % from reading from -90 to -50 °C |
| Resolution | 0.1 °C |
| HUMIDITY SENSOR | |
| Measuring range | 0 to 100 % RH |
| Accuracy | ±1.8 % RH |
| Resolution | 0.1% RH |
| DEW POINT | |
| Measuring range | -90 to +70 °C |
| Accuracy | ±1.5 °C at ambient temperature T <25 °C and RH >30 % |
| Resolution | 0.1 °C |
| GENERAL TECHNICAL DATA | |
| Operating temperature | -30 to +70 °C |
| Channels | internal temperature and humidity sensor, 1x external temperature probe |
| Memory | 500,000 values in noncyclic logging mode; 350,000 values in cyclic record mode |
| Recording interval | adjustable from 1 s to 24 h |
| Display and alarm refresh | adjustable 1 s, 10 s, 1 min |
| Recording mode | noncyclic – data logging stops after filling the memory cyclic – after filling memory oldest data is overwritten by new |
| Real time clock | year, leap year, month, day, hour, minute, second |
| Power | Lithium battery 3.6V, size AA |
| Protection class | IP67 electronics; IP30 sensors |
| Dimensions | 61 x 93 x 32 mm |
| Weight (including batteries) | approx. 120 g |
| Warranty | 3 years |
Tech Specs
Dataloggers / Data Acquisition/Dataloggers for Environmental Variables/Temperature Dataloggers Templ |
|
|---|---|
| Accuracy | TEMPERATURE SENSOR: ±0.4 °C TEMPERATURE SENSOR – EXTERNAL PROBE PT1000: ±0.2 °C (-50 to +100 °C); ±0.2 % from reading from +100 to +260 °C; ±0.4 % from reading from -90 to -50 °C HUMIDITY SENSOR: ± 1.8 % RH DEW POINT: ±1.5 °C at ambient temperature T <25 °C and RH >30 % |
| LCD Screen | Yes |
| Software Included | Yes |
Test Equipment General Attributes |
|
| Warranty | 3 YEARS |
| IP Rating | IP30, IP67 |
| Product Weight | 0.26 LBS |
| Shipping Weight | 0.3 LBS |
| Battery Type | AA |
| Country of Origin | Czech Republic |




Valoraciones
No hay valoraciones aún.